Safety Precautions in the Operating Room

Since safety practices are important to emphasize, this section will cover some of the situations that are potentially hazardous in the operating room and discuss what might be done to eliminate the hazard.

All personnel should know the location of all emergency medications and equipment in the operating room. This includes drugs, cardiac arrest equipment, and resuscitators. All electrical equipment and plugs must be of the explosion-proof type and bear a label stating such. There should be written schedules of inspections and maintenance of all electrical equipment. Navy regulations prohibit the use of explosive anesthetics in the operating room. These regulations, however, do not mean we can lessen our concern for fire and explosion hazards. The surface of all floors in the operating room must provide a path of electrical conductivity between all persons and equipment making contact with the floor to prevent the accumulation of dangerous electrostatic charges. All furniture and equipment should be constructed of metal or of other electrically conductive material and should be equipped with conductive leg tips, casters, or equivalent devices. Periodic inspections should be made of leg tips, tires, casters, or other conductive devices of furniture and equipment. These inspections will ensure that they are maintained free of wax, lint, or other foreign material that may insulate them and defeat the purpose for which they are used. Excess lubrication of casters should be avoided to prevent accumulation of oil on conductive wheels. Dry graphite and graphite oil are the preferred lubricants.

Rubber accessories for anesthesia machines should be of the conductive type, plainly labeled as such, and routinely tested to ensure that conductivity is maintained. It is essential that all replacement items be of conductive material.

All personnel entering the operating room should be in electrical contact with the conductive floor by wearing conductive footwear or an alternative method of providing a path of conductivity. Conductive footwear and other personnel-to-floor conductive equipment should be tested on a regularly scheduled basis.

All apparel worn in the operating room should be made of a nonstatic-producing material. Fabrics of 100 percent cotton are the most acceptable. Fabrics made of synthetic blends may be used only if they have been treated by the manufacturer for use in the operating room. Wool blankets and apparel made of untreated synthetic fabrics are not permitted in the operating room.

Operating rooms must have adequate air-conditioning equipment to maintain relative humidity and temperature within a constant range. The relative humidity should be kept at 55 to 60 percent. This level will reduce the possibility of electrostatic discharge and possible explosion of combustible gases. The temperature should be chosen on the basis of the well-being of the patient. The recommended temperature is between 65 E and 74 E F. The control of bacteria carried on dust particles is facilitated when the recommended humidity and temperature are maintained.

All oxygen cylinders in use or in storage will be tagged with DD Form 1191, Warning Tag for Medical Oxygen Equipment, and measures will be taken to ensure compliance with instructions 1 through 7 printed on the form. Anadditional tag is required on all cylinders to indicate "EMPTY," "IN USE," or "FULL." Safety precautions should be conspicuously posted in all areas in which oxygen cylinders are stored and in which oxygen therapy is being administered. This posting should be made so it will immediately make all personnel aware of the precautionary measures required in the area.

All electrical service equipment, switchboards, or panelboards should be installed in a nonhazardous location. Devices or apparatus that tend to create an arc, sparks, or high temperatures must not be installed in hazardous locations unless these devices are in compliance with the National Electrical Code. Lamps in a fixed position will be enclosed and will be properly protected by substantial metal guards or other means where exposed to breakage. Cords for portable lamps or portable electrical appliances must be continuous and without switches from the appliance to the attachment plug. Such cords must contain an insulated conductor to form a grounding connection between the electrical outlet and the appliance.